DROP · SHOCK · VIBRATION / EQUIPMENT & INSTRUMENTATION

APPLICATION NOTE · T8-DSA-601

Sound Pressure and Sound Quality Analysis

Two sounds can both measure 85 dB, one low and full, the other sharp and harsh. A sound level meter's single number cannot tell them apart, yet product noise complaints often come from sounds that are unpleasant rather than too loud. On one multi-channel acquisition system, this module performs sound level statistics, octave spectra, Zwicker psychoacoustic sound quality and automatic tone detection. It gives 16 indicators per channel, including loudness, sharpness, roughness, fluctuation strength and articulation index, so sound quality becomes engineering data that can be improved and accepted.

Category
Acoustics
Applies to
T8-3X Dynamic Signal Analyzer
Content reviewed

Function overview

The problem it solves

The A-weighted level answers only how loud a sound is. Sounds of equal level — a low hum or a high hiss — are perceived very differently: the first is dull but tolerable, the second irritating. Most noise complaints about appliances, vehicles and motors are not about exceeding a level but about harsh high-frequency content or noticeable modulation — sound quality problems.

The traditional approach uses a sound level meter for levels, an octave analyzer for spectra and separate sound quality software for psychoacoustic indicators: three systems and three data sets that are hard to align in time and cannot analyze many channels at once. Integrated in the T8 vibration and acoustic system, the module computes sound level statistics, octave spectra, FFT spectra and Zwicker sound quality for eight channels at every refresh, answering four questions in one measurement: how loud, what content, how pleasant and whether tones are present.

Functions

FunctionDescription
Multi-channel sound pressure waveformsTime waveforms (Pa) of the selected channels, to confirm signal quality
Statistics table per channelEquivalent level, peak level, sound exposure level, kurtosis, crest factor, impulse factor, form factor, noise dose, TWA, loudness, loudness level, sharpness, roughness, fluctuation strength and articulation index
Octave spectra1/3 octave (Z, A or C weighting), 1/1 octave and 1/12 octave, in the stepped display used by acoustic instruments
Specific loudness spectrumLoudness distributed over the 24 critical bands (Bark) of the ear, showing which bands produce the loudness
Excitation level spectrumExcitation distribution including masking, reflecting what the ear actually hears
FFT spectrumFFT magnitude (dB) and FFT phase views
Fast level historyLevel against time with IEC 61672 Fast time weighting (125 ms), continuous across data blocks
Tone and peak detectionAutomatic search of spectral peaks; components more than 6 dB prominent are marked as audible tones
Data exportCSV report with the statistics and the 1/3-octave table
Stay on top and helpKeep the window in front; built-in reference

Screen layout

  1. Statistics table (top): all 16 indicators for each channel, updated at every refresh.
  2. Time waveform (middle): sound pressure (Pa) of the selected channel against time in ms.
  3. Octave view (lower left): 1/3 octave Z/A/C, 1/1, 1/12, specific loudness and excitation level.
  4. Spectrum view (lower right): FFT magnitude in dB, FFT phase or the Fast (125 ms) level history.

The left panel selects the channels to analyze and the analysis channel used for tone detection; below it, the tone and peak list shows the main peak frequencies, levels and tone marks.

31.5631252505001k2k4k8k305070901/3-octave band (Hz)Sound pressure level (dB)Z-weightedA-weighted
1/3-octave spectrum in the stepped display used by acoustic instruments, with A-weighted levels. Redrawn schematic.

Applications

Sound quality of appliances and electromechanical products

For air conditioners, refrigerators, washing machines, vacuum cleaners, motors and pumps, loudness decides whether a limit is exceeded, sharpness whether the sound is harsh, and roughness or fluctuation strength whether there is annoying modulation. Quantifying these in development, with octave and specific loudness spectra to find the frequency source of harsh components, gives a direct basis for acoustic improvement.

Occupational noise exposure

Workplace assessment concerns not only the instantaneous level but the accumulated exposure over eight hours. The module gives the noise dose (%) and the eight-hour time-weighted average level (TWA) directly for comparison with occupational limits.

Tonality of fans and wind turbines

Fan and wind turbine noise rules (such as IEC 61400-11) require a decision on audible tones, which strongly increase the risk of complaints. The module searches spectral peaks, calculates tone prominence and marks components above 6 dB as tones without manual reading.

Workshop and environmental noise statistics

Fluctuating noise, such as intermittent machines or traffic, is badly described by instantaneous levels. The module calculates statistics over the whole period (maximum and minimum level, sound exposure level SEL) and, with the octave spectrum, identifies the spectral composition and main contributing bands.

Speech communication environments

Control rooms, command centers and cabins need speech intelligibility. The articulation index (AI, 0 to 1) quantifies how clearly speech is transmitted against the current background, guiding noise control targets and public address design.

Training and demonstration

With the supplied example data, all functions — stepped octave spectra, Bark specific loudness, Fast level history and tone detection — can be demonstrated without a sound calibrator.

What each indicator tells you

IndicatorQuestionEngineering meaning
Equivalent level LeqHow loud is the sound overall?Basic quantity for compliance; an energy average not misled by fluctuations
Peak level LpeakHow strong are instantaneous impacts?Evaluation of impact and blast sounds; a high crest factor indicates impacts
Sound exposure level SELWhat is the event energy normalized to 1 second?Fair comparison of events of different duration (for example single pass-by events)
Kurtosis, crest, impulse and form factorsDoes the waveform contain impacts?Dimensionless statistics that identify impulsive noise such as clicks or foreign-object impacts
Noise dose / TWAHow much exposure has a worker accumulated?An exposure estimate only when the configured regime and representative task durations are verified
Loudness N / loudness levelHow loud does it sound?Closer to perception than dB; a direct measure of product loudness
Sharpness SIs the sound harsh?Measure of the high-frequency share; often the first target of noise improvement
Roughness R / fluctuation strengthDoes the sound tremble or rasp?Quantifies modulation noises (gears, blade passing, electromagnetic pulsation)
Articulation index AICan speech be understood in this noise?Design target for communication and public address environments (0–1, higher is clearer)
Octave / specific loudness spectraWhere in frequency is the problem?Locates the problem in frequency
Tone prominenceAre there audible tones?Key factor for regulatory decisions (wind turbines, fans) and subjective annoyance

How it works

Sound pressure level and frequency weighting

Lp = 20 · lg(p / 20 µPa) [dB]

The reference of 20 µPa is the threshold of hearing at 1 kHz. A and C weightings correct the spectrum band by band according to IEC 61672: A-weighting models the ear's low sensitivity at low frequencies (used for environmental and occupational noise); C-weighting is flat and keeps low-frequency energy (used for impulse and peak evaluation).

Equivalent level, exposure level and statistics

The equivalent continuous level Leq is the energy average of the squared sound pressure over the analysis period. The sound exposure level normalizes the total energy to one second, SEL = Leq + 10 · lg(T), for fair comparison of events of different duration. Kurtosis, crest factor (peak / RMS), impulse factor (peak / mean absolute value) and form factor (RMS / mean absolute value) are dimensionless waveform statistics with typical values for sine and random noise; clear deviations indicate impacts or modulation.

Octave spectra (IEC 61260)

The range from 20 Hz to 16 kHz is divided into standard bands: 10 bands for 1/1 octave, 30 bands for 1/3 octave and the corresponding finer division for 1/12 octave. Each band level is the energy sum of the FFT lines in the band:

L_band = 10 · lg(Σ p² / p0²) [dB]

Bands are drawn as steps from lower edge to upper edge, as on acoustic instruments. The 1/3-octave view supports Z, A and C weighting, with corrections at each band center. Empty bands are clamped to 0 dB so the curve does not drop abnormally.

Fast time-weighted level history (IEC 61672)

Env(i) = α · x²(i) + (1 − α) · Env(i − 1), α = 1 − exp(−1 / (0.125 · Fs))

The Fast setting of a sound level meter is a first-order exponential average of the instantaneous power with a 125 ms time constant. The averaging state is kept between data blocks, so long histories have no jump at the start of each block, consistent with a real sound level meter.

Zwicker loudness and critical bands (ISO 532-1)

The cochlea acts like a filter bank: 24 critical bands (Bark scale) each sense the energy in their range. Loudness is calculated in four steps:

StepOperationBasis
1Build the 1/3-octave spectrum from the windowed FFT magnitudeIEC 61260 band division
2Map 1/3-octave energy to 24 Bark critical bandsZwicker frequency-to-Bark transform
3Add the hearing threshold and apply masking (upward masking 25 dB/Bark, downward 65 dB/Bark)ISO 389-7 threshold; psychoacoustic masking
4Calculate specific loudness N′(z) from the excitation of each band, sum to total loudness N (sone) and convert to loudness level (phon)ISO 532-1 Zwicker method

Ln = 40 + 10 · log2 N (for N ≥ 1 sone)

Each doubling of loudness adds 10 phon — the essential difference between perception and physical decibels.

Sharpness, roughness and fluctuation strength

Sharpness is the weighted average of specific loudness over critical band rate; bands above 15.8 Bark are weighted increasingly, emphasizing harsh high-frequency content, in acum. Roughness and fluctuation strength describe rapid variation of the envelope: the signal is rectified and low-pass filtered to extract the envelope, and the modulation depth and frequency are estimated in the 20–200 Hz range (roughness, with maximum sensitivity at 70 Hz modulation) and the 0.5–20 Hz range (fluctuation strength, maximum at 4 Hz), then converted with psychoacoustic weighting curves.

Articulation index (ANSI S3.5)

The 15 one-third-octave bands above 200 Hz are weighted by their importance for speech intelligibility. The signal-to-noise ratio of each band between −15 and +30 dB is mapped linearly to 0–1, and the weighted sum is the articulation index. The closer AI is to 1, the easier speech is to understand.

Tone prominence (IEC 61400-11 basis)

Local maxima in the FFT magnitude spectrum are candidate peaks. For each, the difference between its level and the mean noise level in its critical band (prominence) is calculated. Peaks more than 6 dB prominent are audible tones and marked in the tone list, which also gives the peak frequency (Hz) and level (dB) of up to six main peaks.

Performance

All indicators are calculated per channel at every refresh. The windowed FFT of the main analysis chain is reused rather than computed twice, and data-heavy views such as 1/12 octave are calculated on demand, so continuous refreshing is possible with many channels for online monitoring.

Reference standards

StandardTitleRelevance
IEC 61672-1 / GB/T 3785.1Sound level meters, Part 1: specificationsDesign basis for A, C and Z frequency weighting and Fast (125 ms) time weighting
IEC 61260-1 / GB/T 3241Octave-band and fractional-octave-band filters1/1, 1/3 and 1/12 octave bands and nominal center frequencies
ISO 532-1Acoustics — Methods for calculating loudness — Zwicker methodLoudness (sone), loudness level (phon), specific loudness and the masking model
ISO 389-7Acoustics — Reference zero for the calibration of audiometric equipment — Part 7: threshold of hearing under free-field and diffuse-field conditionsThreshold table for the Bark bands
ANSI S3.5Methods for calculation of the articulation indexBand weights and SNR mapping of the articulation index
IEC 61400-11Wind turbines — Acoustic noise measurement techniquesBasis for the tone prominence decision (6 dB threshold)
OSHA 29 CFR 1910.95Occupational noise exposure (U.S.)Noise dose (8 h at 90 dB = 100 %, 5 dB exchange rate) and TWA calculation

Key capabilities

CapabilityBenefit
Three roles in one systemSound level meter, octave analyzer and sound quality analyzer in one
Eight synchronized channelsSeveral points at once (around the machine, operator position, background) with per-channel comparison
Complete psychoacoustic indicatorsLoudness, sharpness, roughness, fluctuation strength and articulation index for sound quality improvement
Up to 1/12 octaveHigh-resolution band spectra for locating narrow-band noises
Excitation level spectrum with maskingShows not only what is physically present but what the ear hears
Automatic tone decisionAudible tones marked on the IEC 61400-11 basis without manual reading
Direct occupational readingsNoise dose and TWA without manual conversion
Continuous Fast historyLevel history continuous across data blocks, as on a sound level meter
One-step CSV reportStatistics and octave table exported together
Shared platform with vibrationAcoustic and vibration analysis share acquisition hardware and data, supporting combined noise and vibration work

Operating workflow

Preparation

  1. Microphones: 0.5–1 m from the source, away from air outlets and reflecting surfaces; with several channels keep the distance to the source similar.
  2. Calibration: calibrate with a sound calibrator (for example 94 dB at 1 kHz) and check that the equivalent level in the table matches after refresh.
  3. Acquisition: a sampling rate of at least 51.2 kHz for a 20 kHz analysis bandwidth; choose the number of points for the required 1/12-octave low-frequency resolution.

Procedure

StepActionNote
1Select analysis channelsOnly enabled channels can be selected
2Select the analysis channelThe reference channel used for tone detection
3RefreshAll selected channels are calculated and the four areas updated
4Check the statistics tableThe 16 indicators per channel, such as Leq, peak and dose
5Switch the octave view1/3 octave Z/A/C, 1/1, 1/12, specific loudness, excitation level
6Switch the spectrum viewFFT magnitude dB, FFT phase, Fast (125 ms) level history
7Read the tone listPeak frequencies, levels and tone marks
8ExportCSV with statistics and the 1/3-octave table
9Stay on topKeep the window in front for side-by-side work

Recommended views

PurposeViewNote
Sound quality improvementSpecific loudness spectrum with sharpness and loudnessThe peak band of specific loudness is the source band of harsh components
Spectral composition of noise1/3 octave, A-weightedThe A-weighted spectrum matches perceived distribution
Tones and abnormal soundsFFT magnitude with the tone listHigh FFT resolution with tone marks gives a direct conclusion
Modulation noisesRoughness and fluctuation strengthHigh values indicate clear envelope modulation
Workplace noiseDose, TWA and LeqEvaluated for the current measurement period; record long periods in segments
Speech environmentArticulation indexAI below 0.3 indicates insufficient intelligibility; reduce noise or amplify

Worked example: reading the statistics table

Channel 1, near field of a fan prototype:

  • Equivalent level 83.9 dB and sound exposure level 89.1 dB.
  • Kurtosis 1.49 and crest factor 2.31, both typical of steady noise, so the signal contains no clear impacts.
  • No exposure conclusion is drawn from this example: dose/TWA settings, representative measurement periods and daily task durations are not established.
  • Loudness 51.63 sone (about 97 phon) with a high sharpness of 2.872 acum, indicating a sharp sound. The specific loudness and 1/3-octave spectra show strong contributions between 1 and 4 kHz, so improvement should target the air path and blade design in that range.
  • The tone list shows main peaks at 125 Hz, 315 Hz and others; peaks marked as tones are more than 6 dB prominent and their mechanical or electromagnetic sources should be investigated first.

The reading order — overall level (Leq), character (kurtosis, crest factor), perception (loudness, sharpness) and frequency location (octave, specific loudness, tone list) — leads from data to improvement measures in four steps.