APPLICATION NOTE · T8-DSA-405
The same vibration or noise signal leads to different conclusions depending on who or what is “listening”: the ear, the hand and arm, the spine and an instrument each have their own sensitive frequency range. Frequency weighting gives the measuring instrument the matching sensitivity. In one view, this module performs 1/3-octave analysis, switches between nine standard weightings and converts levels to more than a dozen dB references, so one measurement covers acoustics, occupational health, vehicle ride comfort and electronic measurement.
Raw physical data shows the instrument's view, while people, regulations and contracts use a human view. Weighting converts the first into the second, and the module makes the switch a single selection.
A weighted spectrum describes the measured vibration or sound during the selected analysis period. Occupational exposure assessment also requires representative sampling, exposure duration, the required axis combination and the applicable regulation. A current-frame weighted RMS or dB value alone does not establish daily exposure or compliance.
| Function | Description |
|---|---|
| 1/3-octave spectrum | 47 standard bands from 0.5 Hz to 20 kHz, shown as bars or as a contour line |
| Nine weightings | Unweighted, A, C, hand-arm, whole-body Z / XY / M, ACGIH XY / Z |
| 19 reference values | Sound pressure, sound intensity, sound power, micro-vibration, velocity, displacement, force, voltage and power; dB values converted automatically |
| Total RMS and total level | Weighted total RMS and total weighted dB shown together as inputs to the selected assessment method |
| Exponential averaging | 1–1000 averages for stable readings; reset at any time |
| Eight synchronized channels | Each channel averages independently |
| Two views | Time waveform above, weighted spectrum below |
| Unit consistency check | A warning when the 0 dB reference does not match the channel unit, preventing wrong dB values |
| Built-in reference | Acoustic calibration check points, occupational limit tables and typical reference values |
| Stay on top and maximize | Keep the window in front; double-click a plot to maximize it |
Handle vibration of drills, angle grinders, breakers and rock drills is weighted with Wh. The combined weighted vibration magnitude ahv is an input to the daily exposure calculation A(8), together with the duration of each task. Compare the duration-normalized result with the applicable occupational regulation; ISO 5349 is a measurement and evaluation method, not a universal exposure limit.
Seat vibration can be evaluated using whole-body Wk (vertical) or Wd (horizontal) weighting. Comfort and occupational health assessments use different criteria. Daily exposure assessment also requires exposure duration and the axis factors and method specified for the task; the current weighted RMS alone is insufficient.
A-weighted sound pressure is used in many product and environmental specifications. For example, a hypothetical refrigerator contract might specify 45 dBA under stated test conditions; that is not a universal product limit. C-weighting retains more low-frequency energy and may be used for peak assessments when required. With the 20 µPa reference, the module displays dBA or dBC.
A sound calibrator at 1 kHz and 94 dB checks the measurement chain at that frequency. A- and C-weighting are nominally 0 dB at 1 kHz, so the three readings should be close to the calibrator value within the applicable tolerances. This single-point check does not validate full-band response or instrument conformity.
Vibration on precision instrument foundations or lithography platforms can be below 1 µm/s², where normal unit readings are long strings of zeros. Converted to dB with a 1 µm/s² (or 0.1 µg) reference, the reading becomes a two-digit number; velocity (1 nm/s) and displacement references are also available. References for underwater sound (1 µPa) and for sound intensity and sound power (ISO 1683) are included.
With the dBV (1 V), dBu (0.775 V) or dBm (1 mW) reference, the same spectrum engine analyzes audio equipment levels and power in frequency bands.
| Measurement | Question it answers | Engineering use |
|---|---|---|
| Unweighted 1/3-octave spectrum | In which bands the energy physically lies | Locating fault bands and identifying noise or vibration sources |
| A-weighted dBA | How loud it sounds to the human ear | Statutory basis for product, environmental and occupational noise |
| C-weighted dBC | Total energy including low-frequency and impulse content | Impulse peak checks and low-frequency noise |
| Hand-arm weighted ahv | What is the weighted vibration magnitude at the tool handle? | Input to an ISO 5349 daily exposure assessment with measured task durations and the applicable occupational limits |
| Whole-body weighted aw (Wk / Wd) | What is the weighted seat vibration magnitude? | Input to comfort or health assessment under the selected method and exposure duration |
| ACGIH weighting | How does the weighted magnitude relate to the selected occupational method? | Select the applicable occupational assessment reference, edition and exposure method |
| Velocity or displacement level in dB | How small a micro-vibration is | Concise expression for precision foundations and optical tables |
| Total RMS and total level | The combined value over all bands | A measured magnitude; the assessment also needs the criterion, duration and measurement conditions |
The calculation has four steps: divide into bands, sum energy, apply the standard weighting, and combine the total level.
Each frame is windowed with a Hanning window and transformed by FFT. Windowing loses energy, so the software applies the amplitude correction (× 2) and the energy correction (× √(8/3)), so band energies match the true values. If this step is wrong, all later dB values are systematically low.
The 1/3-octave scale doubles the frequency every three bands; each band is bounded by its center frequency multiplied or divided by 2^(1/6). By Parseval's theorem, the band RMS is the square root of the sum of squared RMS values of all FFT lines in the band. The peak spectrum is divided by √2 to give RMS, and the band edges are limited to the actual analysis range (resolution to upper analysis frequency). This also explains why white noise rises by 3 dB per octave in an octave spectrum while pink noise is flat: higher bands contain more lines.
Each weighting curve is a table of dB corrections per band defined by the standard; the software multiplies each band by the linear factor 10^(dB/20).
| Weighting | Standard | Meaning of the curve |
|---|---|---|
| A | IEC 61672 | Approximates the ear's sensitivity at a moderate loudness of 40 phon: strong attenuation at low frequencies, 0 dB at 1 kHz and roll-off at high frequencies |
| C | IEC 61672 | Approximates a high loudness of 100 phon; almost flat from 20 Hz to 1 kHz, so total energy is kept |
| Hand-arm Wh | ISO 5349-1 | Frequency weighting for hand-transmitted vibration; components below 6.3 Hz and above about 1.6 kHz are excluded |
| Whole-body Z (Wk) | ISO 2631-1 | Weights vertical vibration around the 4–8 Hz range where the spine and abdomen resonate |
| Whole-body XY (Wd) | ISO 2631-1 | The balance system is most sensitive to horizontal motion around 1–2 Hz |
| Building vibration (Wm) | ISO 2631-2 | Weighting for vibration in buildings; not a substitute for a whole-body occupational exposure assessment |
| ACGIH XY / Z | ACGIH TLV | North American occupational limit system, aimed at long-term injury limits, with more weight at low frequencies than ISO |
When a weighting is selected, the displayed band range changes accordingly (for example 10 Hz–16 kHz for A and C, 6.3 Hz–1.6 kHz for hand-arm and 0.5–80 Hz for whole-body) and components outside it are not counted.
Total RMS = √(Σ weighted band RMS²) Total level = 20 · lg(total RMS / 0 dB reference)
In contour mode the total level is the energy sum of the band levels, 10 · lg Σ 10^(band dB / 10); the two methods are equivalent and cross-check each other. The 0 dB references follow international conventions: 20 µPa for airborne sound (threshold of hearing), 1 µm/s² for micro-vibration and 1 nm/s for velocity. A dB value is relative and has meaning only with its reference.
Spectra fluctuate from frame to frame during live measurement. Exponential averaging (new data weighted 1/N into the history) converges quickly while still following changes. The number of averages is 1–1000; restart the average after changing the operating condition.
| Standard | Title | Relevance |
|---|---|---|
| IEC 61672 | Sound level meters (including A and C frequency weighting) | A and C weighting tables and the 20 µPa airborne sound reference |
| IEC 61260 | Octave-band and fractional-octave-band filters | 1/3-octave band division and center frequencies |
| ISO 1683 | Acoustics: preferred reference values for acoustical and vibratory levels | dB references for underwater sound, sound intensity, sound power and vibration |
| ISO 5349-1 | Hand-transmitted vibration: measurement and evaluation | Wh weighting and measurement input for duration-normalized hand-arm exposure |
| ISO 2631-1 | Whole-body vibration: evaluation of human exposure | Wk and Wd weighting; select the method and duration required for the assessment |
| ACGIH TLV | Threshold limit values for occupational exposure (whole-body vibration) | ACGIH XY / Z weighting |
| GBZ 2.2 | Occupational exposure limits for hazardous agents in the workplace: physical agents (noise, vibration) | Occupational health limits in China |
| GB 12348 | Emission standard for industrial enterprise noise at the boundary | Boundary dBA limits in China |
| Capability | Benefit |
|---|---|
| Nine weightings in one instrument | Functions of a sound level meter, a human vibration meter and an environmental noise meter in one system |
| 19 dB references converted automatically | For acoustic, vibration and electrical work without reference tables |
| Energy-correct window correction | Band energies are restored correctly, so dB values pass calibrator checks |
| Direct total level for limit checks | Weighted levels and magnitudes support assessment when combined with the required time and measurement basis |
| Unit consistency check | Mismatched reference and channel units are flagged before a wrong dB value is reported |
| Built-in check points and limit tables | 94 dB calibration checks and ISO/GBZ limits at hand |
| Eight channels with independent averaging | Several measurement points in one setup |
| Fast exponential averaging | Stable field readings; one-step reset after a change of condition |
| Bar and contour display | Bars for band comparison, contour for spectral shape |
| Shared data with other modules | The same data can be analyzed further with FFT, sound quality and octave functions |
| Step | Action | Notes |
|---|---|---|
| 1 | Select measurement channels | Each check box shows the channel's engineering unit; select only the channels in use |
| 2 | Select the weighting | By purpose: A or C for noise, hand-arm for hand-held tools, whole-body Z or XY for seats, ACGIH for North American projects |
| 3 | Select the 0 dB reference | The reference unit must match the channel unit (the software warns if it does not) |
| 4 | Set averaging | Usually 10; increase if readings fluctuate; restart after a change of condition |
| 5 | Select the display mode | Bars to compare bands; contour to see the spectral shape and the total level |
| 6 | Read and evaluate | Record the weighted RMS, level, axes, reference, measurement period and exposure duration; evaluate under the applicable method |
| Task | Weighting | 0 dB reference | Limit reference |
|---|---|---|---|
| Product or environmental noise | A | Airborne sound 20 µPa | Select the applicable product or environmental criterion and required averaging period |
| Impulse noise peak | C | Airborne sound 20 µPa | Select the applicable peak-noise criterion and confirm the peak detector and weighting |
| Hand-held tool exposure | Hand-arm Wh | 1 µm/s² or 0.1 µg | Calculate daily A(8) using measured ahv and task duration; use the applicable regulation |
| Seat whole-body vibration | Whole-body Z (vertical) / XY (horizontal) | 1 µm/s² or 0.1 µg | Apply the required axis factors, exposure duration and health or comfort assessment method |
| Micro-vibration | Unweighted with the specified vibration criterion | 1 µm/s², 0.1 µg or 1 nm/s | Permissible vibration curve of the precision equipment |
| Audio levels | Unweighted | dBV, dBu or dBm | Level design specification |
Measured data on channel 2, read in four steps: total level, spectral shape, bands, conclusion.