Weighted 1/3-Octave Analysis for Sound and Human Vibration

The same vibration or noise signal leads to different conclusions depending on who or what is “listening”: the ear, the hand and arm, the spine and an instrument each have their own sensitive frequency range. Frequency weighting gives the measuring instrument the matching sensitivity. In one view, this module performs 1/3-octave analysis, switches between nine standard weightings and converts levels to more than a dozen dB references, so one measurement covers acoustics, occupational health, vehicle ride comfort and electronic measurement.

Category
Environmental and human vibration
Applies to
T8-3X Dynamic Signal Analyzer
Content reviewed

Function overview

Raw physical data shows the instrument's view, while people, regulations and contracts use a human view. Weighting converts the first into the second, and the module makes the switch a single selection.

The problem it solves

A weighted spectrum describes the measured vibration or sound during the selected analysis period. Occupational exposure assessment also requires representative sampling, exposure duration, the required axis combination and the applicable regulation. A current-frame weighted RMS or dB value alone does not establish daily exposure or compliance.

Functions

FunctionDescription
1/3-octave spectrum47 standard bands from 0.5 Hz to 20 kHz, shown as bars or as a contour line
Nine weightingsUnweighted, A, C, hand-arm, whole-body Z / XY / M, ACGIH XY / Z
19 reference valuesSound pressure, sound intensity, sound power, micro-vibration, velocity, displacement, force, voltage and power; dB values converted automatically
Total RMS and total levelWeighted total RMS and total weighted dB shown together as inputs to the selected assessment method
Exponential averaging1–1000 averages for stable readings; reset at any time
Eight synchronized channelsEach channel averages independently
Two viewsTime waveform above, weighted spectrum below
Unit consistency checkA warning when the 0 dB reference does not match the channel unit, preventing wrong dB values
Built-in referenceAcoustic calibration check points, occupational limit tables and typical reference values
Stay on top and maximizeKeep the window in front; double-click a plot to maximize it
101001k10k-60-40-200Frequency (Hz)Weighting (dB)1 kHz: 0 dBA-weightingC-weighting
A- and C-frequency weighting curves calculated from the IEC 61672 definitions; both are 0 dB at 1 kHz.

Applications

Hand-held tool vibration exposure (ISO 5349)

Handle vibration of drills, angle grinders, breakers and rock drills is weighted with Wh. The combined weighted vibration magnitude ahv is an input to the daily exposure calculation A(8), together with the duration of each task. Compare the duration-normalized result with the applicable occupational regulation; ISO 5349 is a measurement and evaluation method, not a universal exposure limit.

Ride comfort of vehicles and machinery (ISO 2631)

Seat vibration can be evaluated using whole-body Wk (vertical) or Wd (horizontal) weighting. Comfort and occupational health assessments use different criteria. Daily exposure assessment also requires exposure duration and the axis factors and method specified for the task; the current weighted RMS alone is insufficient.

Product noise and compliance (dBA)

A-weighted sound pressure is used in many product and environmental specifications. For example, a hypothetical refrigerator contract might specify 45 dBA under stated test conditions; that is not a universal product limit. C-weighting retains more low-frequency energy and may be used for peak assessments when required. With the 20 µPa reference, the module displays dBA or dBC.

Acoustic calibration and measurement chain check

A sound calibrator at 1 kHz and 94 dB checks the measurement chain at that frequency. A- and C-weighting are nominally 0 dB at 1 kHz, so the three readings should be close to the calibrator value within the applicable tolerances. This single-point check does not validate full-band response or instrument conformity.

Micro-vibration and special measurements

Vibration on precision instrument foundations or lithography platforms can be below 1 µm/s², where normal unit readings are long strings of zeros. Converted to dB with a 1 µm/s² (or 0.1 µg) reference, the reading becomes a two-digit number; velocity (1 nm/s) and displacement references are also available. References for underwater sound (1 µPa) and for sound intensity and sound power (ISO 1683) are included.

Electronics and audio engineering

With the dBV (1 V), dBu (0.775 V) or dBm (1 mW) reference, the same spectrum engine analyzes audio equipment levels and power in frequency bands.

What each value answers

MeasurementQuestion it answersEngineering use
Unweighted 1/3-octave spectrumIn which bands the energy physically liesLocating fault bands and identifying noise or vibration sources
A-weighted dBAHow loud it sounds to the human earStatutory basis for product, environmental and occupational noise
C-weighted dBCTotal energy including low-frequency and impulse contentImpulse peak checks and low-frequency noise
Hand-arm weighted ahvWhat is the weighted vibration magnitude at the tool handle?Input to an ISO 5349 daily exposure assessment with measured task durations and the applicable occupational limits
Whole-body weighted aw (Wk / Wd)What is the weighted seat vibration magnitude?Input to comfort or health assessment under the selected method and exposure duration
ACGIH weightingHow does the weighted magnitude relate to the selected occupational method?Select the applicable occupational assessment reference, edition and exposure method
Velocity or displacement level in dBHow small a micro-vibration isConcise expression for precision foundations and optical tables
Total RMS and total levelThe combined value over all bandsA measured magnitude; the assessment also needs the criterion, duration and measurement conditions

How it works

The calculation has four steps: divide into bands, sum energy, apply the standard weighting, and combine the total level.

Step 1: FFT spectrum and window correction

Each frame is windowed with a Hanning window and transformed by FFT. Windowing loses energy, so the software applies the amplitude correction (× 2) and the energy correction (× √(8/3)), so band energies match the true values. If this step is wrong, all later dB values are systematically low.

Step 2: 1/3-octave band energy

The 1/3-octave scale doubles the frequency every three bands; each band is bounded by its center frequency multiplied or divided by 2^(1/6). By Parseval's theorem, the band RMS is the square root of the sum of squared RMS values of all FFT lines in the band. The peak spectrum is divided by √2 to give RMS, and the band edges are limited to the actual analysis range (resolution to upper analysis frequency). This also explains why white noise rises by 3 dB per octave in an octave spectrum while pink noise is flat: higher bands contain more lines.

Step 3: band-by-band weighting

Each weighting curve is a table of dB corrections per band defined by the standard; the software multiplies each band by the linear factor 10^(dB/20).

WeightingStandardMeaning of the curve
AIEC 61672Approximates the ear's sensitivity at a moderate loudness of 40 phon: strong attenuation at low frequencies, 0 dB at 1 kHz and roll-off at high frequencies
CIEC 61672Approximates a high loudness of 100 phon; almost flat from 20 Hz to 1 kHz, so total energy is kept
Hand-arm WhISO 5349-1Frequency weighting for hand-transmitted vibration; components below 6.3 Hz and above about 1.6 kHz are excluded
Whole-body Z (Wk)ISO 2631-1Weights vertical vibration around the 4–8 Hz range where the spine and abdomen resonate
Whole-body XY (Wd)ISO 2631-1The balance system is most sensitive to horizontal motion around 1–2 Hz
Building vibration (Wm)ISO 2631-2Weighting for vibration in buildings; not a substitute for a whole-body occupational exposure assessment
ACGIH XY / ZACGIH TLVNorth American occupational limit system, aimed at long-term injury limits, with more weight at low frequencies than ISO

When a weighting is selected, the displayed band range changes accordingly (for example 10 Hz–16 kHz for A and C, 6.3 Hz–1.6 kHz for hand-arm and 0.5–80 Hz for whole-body) and components outside it are not counted.

Step 4: total RMS and total weighted level

Total RMS = √(Σ weighted band RMS²) Total level = 20 · lg(total RMS / 0 dB reference)

In contour mode the total level is the energy sum of the band levels, 10 · lg Σ 10^(band dB / 10); the two methods are equivalent and cross-check each other. The 0 dB references follow international conventions: 20 µPa for airborne sound (threshold of hearing), 1 µm/s² for micro-vibration and 1 nm/s for velocity. A dB value is relative and has meaning only with its reference.

Stable readings: exponential averaging

Spectra fluctuate from frame to frame during live measurement. Exponential averaging (new data weighted 1/N into the history) converges quickly while still following changes. The number of averages is 1–1000; restart the average after changing the operating condition.

Reference standards

StandardTitleRelevance
IEC 61672Sound level meters (including A and C frequency weighting)A and C weighting tables and the 20 µPa airborne sound reference
IEC 61260Octave-band and fractional-octave-band filters1/3-octave band division and center frequencies
ISO 1683Acoustics: preferred reference values for acoustical and vibratory levelsdB references for underwater sound, sound intensity, sound power and vibration
ISO 5349-1Hand-transmitted vibration: measurement and evaluationWh weighting and measurement input for duration-normalized hand-arm exposure
ISO 2631-1Whole-body vibration: evaluation of human exposureWk and Wd weighting; select the method and duration required for the assessment
ACGIH TLVThreshold limit values for occupational exposure (whole-body vibration)ACGIH XY / Z weighting
GBZ 2.2Occupational exposure limits for hazardous agents in the workplace: physical agents (noise, vibration)Occupational health limits in China
GB 12348Emission standard for industrial enterprise noise at the boundaryBoundary dBA limits in China

Key capabilities

CapabilityBenefit
Nine weightings in one instrumentFunctions of a sound level meter, a human vibration meter and an environmental noise meter in one system
19 dB references converted automaticallyFor acoustic, vibration and electrical work without reference tables
Energy-correct window correctionBand energies are restored correctly, so dB values pass calibrator checks
Direct total level for limit checksWeighted levels and magnitudes support assessment when combined with the required time and measurement basis
Unit consistency checkMismatched reference and channel units are flagged before a wrong dB value is reported
Built-in check points and limit tables94 dB calibration checks and ISO/GBZ limits at hand
Eight channels with independent averagingSeveral measurement points in one setup
Fast exponential averagingStable field readings; one-step reset after a change of condition
Bar and contour displayBars for band comparison, contour for spectral shape
Shared data with other modulesThe same data can be analyzed further with FFT, sound quality and octave functions

Operating workflow

StepActionNotes
1Select measurement channelsEach check box shows the channel's engineering unit; select only the channels in use
2Select the weightingBy purpose: A or C for noise, hand-arm for hand-held tools, whole-body Z or XY for seats, ACGIH for North American projects
3Select the 0 dB referenceThe reference unit must match the channel unit (the software warns if it does not)
4Set averagingUsually 10; increase if readings fluctuate; restart after a change of condition
5Select the display modeBars to compare bands; contour to see the spectral shape and the total level
6Read and evaluateRecord the weighted RMS, level, axes, reference, measurement period and exposure duration; evaluate under the applicable method

Quick reference: weighting and reference

TaskWeighting0 dB referenceLimit reference
Product or environmental noiseAAirborne sound 20 µPaSelect the applicable product or environmental criterion and required averaging period
Impulse noise peakCAirborne sound 20 µPaSelect the applicable peak-noise criterion and confirm the peak detector and weighting
Hand-held tool exposureHand-arm Wh1 µm/s² or 0.1 µgCalculate daily A(8) using measured ahv and task duration; use the applicable regulation
Seat whole-body vibrationWhole-body Z (vertical) / XY (horizontal)1 µm/s² or 0.1 µgApply the required axis factors, exposure duration and health or comfort assessment method
Micro-vibrationUnweighted with the specified vibration criterion1 µm/s², 0.1 µg or 1 nm/sPermissible vibration curve of the precision equipment
Audio levelsUnweighteddBV, dBu or dBmLevel design specification

Worked example: reading a weighted spectrum

Measured data on channel 2, read in four steps: total level, spectral shape, bands, conclusion.

  1. Total level first: the example display reads 137.86 dB. Its interpretation requires the selected weighting, reference, physical unit and analysis period; the number alone cannot establish exposure or compliance.
  2. Spectral shape: the bands from 31.5 to 200 Hz are clearly higher than the rest (about 128–130 dB), so the energy is concentrated at low and mid frequencies. From 315 Hz the bands fall and stay at about 57–60 dB, so high-frequency content is weak. This points to the nature of the source, such as unbalance, resonance or gear mesh.
  3. Compare weightings: with A-weighting the low-frequency bands are reduced strongly and the total level drops noticeably — the difference between physically loud and loud to the ear. With hand-arm weighting for a hand-held tool, the low- and mid-frequency bands are reduced far less than with A-weighting, and they dominate the total.
  4. Conclusion: identify the dominant bands and record the weighted magnitude, reference and measurement conditions. For an exposure assessment, include the task durations and required normalization before comparison with the applicable criterion. This illustrative spectrum does not establish a health pass or fail.